| FORCE
tension
and compression |
Load
cell
Dynamometers
|
1N ¸ 50kN
(0,1kg ¸ 5t) |
0.01% |
OIML
R60 |
10N
¸ 50kN
(1kg ¸
5t) |
0.01% |
| ISO 376 |
(00, 0.5, 1, 2) |
| ASTM E74 |
(AA, A) |
|
10
¸ 1000kN
(1t ¸ 100t) |
0.05%
500 ppm |
| ISO 376 |
(1, 2) |
| ASTM E74 |
(A) |
|
| Quantity |
instrument |
Range |
*Uncertainty |
| RELATIVE
PRESSURE |
Pressure
gauge
Pressure transducers
Pressure
transmitters |
VACUUM:
-1,4kPa ¸ -100kPa
(-1bar) |
(0.64
+ 7,8 x
10-5
p) Pa |
Gas:
1,4kPa ¸ 180kPa
(1,8bar) |
(0,64
+ 7,8 x
10-5
p) Pa |
Gas:
0.014MPa ¸ 7
MPa (70bar) |
(260
+ 5,1 x
10-5
p) Pa |
Oil:
0.4MPa ¸
160MPa (1600bar) |
(250
+ 6,6 x
10-5
p) Pa |
Oil:
160,1MPa ¸
200MPa (2000bar) |
1
x
10-4
p |
p is the pressure in pascal.
| Quantity |
instrument |
Range |
*Uncertainty |
Reference
Standards |
| TORQUE |
Torque
transducers |
1 ¸
2000 Nm |
0.01% |
EA
10-14 |
| Indicating
torque tools and setting torque tools |
1 ¸
2000 Nm |
0.50% |
UNI
EN 26789 |